| Technical Specifications and
Comparison |
2010 |
2020 |
| Photometric Method: Transmission Photometer |
 |
 |
| Light Source: Tungsten Halogen Lamp |
 |
 |
| Photodetector: Silicon Photodiode |
 |
 |
| Wavelength Range: 400 - 750nm |
 |
 |
| Resolution: 0.001 Abs |
 |
 |
| Absorbance Range: 0.000 - 3.300 Abs |
 |
 |
| Accuracy: 2% Abs at 1 Abs, 492nm |
 |
 |
| Linearity: <±0.75% and ±0.005 Abs from 0.1 Abs to 2.5 Abs |
 |
 |
| Reproducibility: <±0.3% at 1 Abs |
 |
 |
| Reading Speed: 96 well - 30 seconds |
 |
 |
| Standard Filters: 405, 450, 492, 620nm |
 |
 |
| Additional filters: on request, filter wheel for up to 8
filters |
 |
 |
| Voltage Range: 90 - 130 and 180 - 250V (autosensing) |
 |
 |
| Frequency range: 47 - 63Hz |
 |
 |
| Serial interface: 9-pin, (RS232) |
 |
 |
| Parallel printer interface: 25-pin |
|
 |
| Standard Software: ADAP Basic |
 |
 |
| Onboard Software |
|
 |
| Display & Keyboard: 25x42 characters color graphics (640x480
pixel) |
|
 |
| Self control: auto-calibration, auto lamp-adjustment, status
report |
 |
 |
| Printer: DOS compatible printers with parallel interface |
|
 |
| Dimensions and Weight: |
|
|
| 32.6cm x 17.3cm x 43.5cm (W x H x D) 6.6Kg weight |
 |
|
| 34.4cm x 25.5cm x 43.0cm (W x H x D) 10Kg weight |
|
 |
| Scope of Supply: |
|
|
| 4 Standard Filters, Dustcover, Spare Fuses, Power Cable,
Serial Cable |
 |
 |
| Standard Software and User Manual on CD |
 |
 |
| Options: |
|
|
| Additional Software ADAP Plus (including ADAP Basic) |
 |
 |
| Additional Software ADAP Expert (including ADAP Plus) |
 |
 |